AUSTIN AIR ANALYSIS

The comparison is listed below:

 
Austin Air
Care 2000
Total unit weight in pounds
45
50
Unit size in inches
23.5 x 14.5 sq
29 x 14 x 15
HEPA filter square foot
80
80
Speed Control
3 speed switch
variable
Filter monitoring
none
pressure guage
UVGI monitoring
none
LED light
Ease of filter replacement
complete unit disassembly
simple door
Carbon weight in pounds
13-15
15
Noise level in decibels
98 high
55 high
 
58 low
43 low
UVGI lamp wattage
none
2 @ 9 w
Lamp germicidal irradiation
none
64 m w-s/cm2 with 2 lamps
UV effective chamber size
0
14" x 15" x 10"

Micro watt sec/cm squared of effective irradiation power used for killing airborne pathogens

0
12,456

This unit has NO UV lamps. The CARE irradiation would be 43,600 m w-s/cm2 at low speed.

The airflow pattern set up with the supply air, out of the side, is poor. Especially on low speed, short cycling of clean air is prevalent. Clean air must be directed into the breathing zone. This way people would be breathing the cleanest air possible, all the time. The CARE 2000 accomplishes this by sending air to the ceiling so as to wash the entire room with clean air and force the dirty air back to the unit along the floor.

The airflow rate is listed at 400 cfm. The measured airflow is actually 320 cfm. The unit would not typically be used on high speed because of the noise; one actually could not carry on a conversation. Medium speed is much lower at 150 cfm.

The "low speed" setting is so low that no real, effective air changes are made and at minimal air cleaning. CFM is tested at 75 cfm.

Filter changing is not indicated with any type of monitor. The CARE is by actual pressure drop across the filter.

The mfg advertises a high efficiency split capacitor motor, which is really a standard 3-speed type motor.

Back to About Us