How and when are the hospital grade NQ unit filters changed?

NQ UNIT CLEANING

The mobile units require proper cleaning at three specific times.
1 - When the unit is moved from room to room
2 - When the unit is serviced
3 - When the unit is stored of left unattended for more than 1 month

The spread of dust, germs and any possible infection must be contained, therefore the cleaning of the NQ units should follow hospital Standard Operating Procedures ( SOP ). If an SOP for mobile units is not in place at your hospital, clinic or building, please follow the following guidelines. These guidelines are offered as suggestions only and do not constitute a guarantee to not spread infection.

Personnel must wear proper masks and gloves before starting. Area should be under negative pressure if at all possible.
1 - remove unit filters; bag and dispose of old filters as medical waste.
2 - wipe down, with standard disinfectant, the inside and outside of the cabinet, this should be done inside the room before transporting.
3 - install new unit filters.
4 - wipe down the UVGI lamps with a damp cloth, do not use any water or detergent on the glass.
5 - do not turn the unit back on until it is located in the new space.

If the unit is not able to be worked on inside the room a complete sealing of the unit from ambient air ( plastic wrap or equal ) should be discussed with the infectious disease department or other such department depending on the hospital staffing.

The bottom opening must get special attention as the wheels are further out then the actual air opening and special consideration is required so as not to have air leaks when transporting. The unit would then be taken to a secure area and the above 5 steps performed.

previous faqback to faqs listnext faq